Difference between revisions of "OR Master Station - 1076036600"
From Zenitel Wiki
(New page: thumb|250px|OR Master Station 107 6036 600 == Features == * The station gives loud-speaking hands-free duplex communication from any location in the room :: - no...) |
(→Specifications) |
||
Line 24: | Line 24: | ||
|Call buttons || 10 Digit buttons, 0-9 <br/> M (manual) and C (cancel) | |Call buttons || 10 Digit buttons, 0-9 <br/> M (manual) and C (cancel) | ||
|- | |- | ||
− | |Mounting || Flush mount backbox, [[ | + | |Mounting || Flush mount backbox, [[Back box - 1000602000|1000602000]] <br/> Surface mount back box, [[Back box - 1000603000|1000603000]] |
|- | |- | ||
|Weight <br/> || 23 kg | |Weight <br/> || 23 kg |
Revision as of 09:05, 5 August 2010
Features
- The station gives loud-speaking hands-free duplex communication from any location in the room
- - no one needs to leave their duty and all present staff can follow the conversation
- High quality intercom station designed to meet the standards of demanding hospital environments
- Excellent sound quality with 3” loudspeaker and ultra-sensitive electret microphone
- Hygienic, easily to clean and sterilize white Mylar covered faceplate
- Splash protected when flush mounted
- Fully equipped master station with access to all STENTOFON AlphaCom XE features
Description
The Operation Room Master Station can be flush or surface mounted using tailored back boxes. The station is designed for ease of operation and excellent voice quality. The front is designed for use in hospital operating theaters and other clean rooms, the Mylar cover eliminates crevices and allows easy wipe-down sterilization. The surface tolerates all common washing and sterilization detergents.
Short call setup times allow instant contact with for instance labs, radiology and doctors for urgent consultation. The station will operate on all STENTOFON AlphaCom XE systems equipped with ASLT cards and gives users the quality and reliability required in critical situations. The STENTOFON AlphaCom XE is an advanced system with a high level of flexibility, audio quality, system management and supervision. Integration with other security and communication systems like access control, alarms, CC TV, public address and telephone systems is a key feature.
Specifications
Size (WxHxD) | 125 x 280 x 66 mm |
Face plate | Aluminium covered with Bonded Mylar Neoprene gasket against back box |
Call buttons | 10 Digit buttons, 0-9 M (manual) and C (cancel) |
Mounting | Flush mount backbox, 1000602000 Surface mount back box, 1000603000 |
Weight |
23 kg |
Temperature | -20°C to +55°C |
Internal Power Amplifier | 10 W |
Power amp operating voltage | 24-30 VDC |
Loudspeaker | External 8 or 20 ohms loudspeaker must be installed |
Material | Cabinet in laquered stel, light beige |
Keypad | 12 buttons: 0-9, M and C |
Cable requirement | 2 pairs for audio plus 1 pair for 24-30 VDC to power amplifier |
Max loudspeaker output |
5 W with 20 ohms loudspeaker 10 W with 8 ohms loudspeaker |
Distance to exchange for audio |
4.0 km (18 AWG - 0.9 mm) 3.4 km (20 AWG - 0.8 mm) 2.0 km (22 AWG - 0.6 mm) |
Distance to power supply |
8 ohms loudspeaker, 24 VDC: 0.6 mm: 70 m 0.8 mm: 100 m 1 mm: 200 m 8 ohms loudspeaker, 30 VDC: 0.6 mm: 150 m 0.8 mm: 250 m 1 mm: 450 m 20 ohms loudspeaker, 24 VDC: 0.6 mm: 150 m 0.8 mm: 250 m 1 mm: 400 m 20 ohms loudspeaker, 30 VDC: 0.6 mm: 350 m 0.8 mm: 650 m 1 mm: 1000 m |
Control of external call device | Potential free relay contact (3A, 30 VDC) |
Connector type | Screw terminals |
Compliance | CE, FCC Part 15 |